Abstract
Sevenfold improved axial resolution has been achieved in three-dimensional widefield fluorescence microscopy, using a novel interferometric technique in which the sample is observed and/or illuminated from both sides simultaneously using two opposing objective lenses. Separate interference effects in the excitation light and the emitted light give access to higher resolution axial information about the sample than can be reached by conventional widefield or confocal microscopes. Here we report the experimental verification of this resolution performance on complex biological samples.
References
Cited by 413 citations (total)
Journal Article Research Support, Non-U.S. Gov't
Fluorescence microscopy with diffraction resolution barrier broken by stimulated emission.
Electronic publication date August 29, 2000
T A Klar [...] S W Hell et al.
The diffraction barrier responsible for a finite focal spot size and limited resolution in far-field fluorescence microscopy has been fundamentally...
Journal Article Research Support, Non-U.S. Gov't Review
Fluorescence interferometry: principles and applications in biology.
Electronic publication date July 30, 2008
Alberto Bilenca [...] Guillermo Tearney et al.
The use of fluorescence radiation is of fundamental importance for tackling measurement problems in the life sciences, with recent demonstrations...