View on Pubmed.gov 11716240 | View on doi.org 10.1253/jcj.65....
Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias.
- First Department of Internal Medicine, Niigata University School of Medicine, Japan. washi@med.niigata-u.ac.jp
Abstract
The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57+/-16 years). During the follow-up period of 27+/-17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with beta-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted.
References
Cited by 28 citations (total)
Journal Article Research Support, Non-U.S. Gov't
Recurrent cardiac events in patients with idiopathic ventricular fibrillation, excluding patients with the Brugada syndrome.
Electronic publication date March 15, 2006
Jean Champagne [...] Pedro Brugada et al.
BACKGROUND
The recurrence of cardiac events in patients with idiopathic ventricular fibrillation (VF) excluding patients with the Brugada syndrome is unclear...
Case Reports Journal Article
T wave oversensing in implantable cardioverter defibrillators.
Electronic publication date August 19, 2006
Steve S Hsu [...] Florin T Deger et al.
The implantable cardioverter defibrillator is the treatment of choice for patients with ventricular tachycardia, especially in the setting of structural...