iDXA, Prodigy, and DPXL dual-energy X-ray absorptiometry whole-body scans: a cross-calibration study.
- Department of Medicine, Obesity Research Center, St. Luke's-Roosevelt Hospital, and Institute of Human Nutrition, Columbia University, New York, NY 10025, USA.
Abstract
PURPOSE
Total body fat, lean, and bone mineral content (BMC) in addition to regional fat and lean mass values for arms, legs, and trunk were compared across a pencil-beam (Lunar DPXL) and 2 fan-beam (GE Lunar Prodigy and GE Lunar iDXA) dual-energy X-ray absorptiometry (DXA) systems.
METHODS
Subjects were a multiethnic sample of 99 healthy adult males (47%) and females (mean+/-SD: age, 46.3+/-16.9 yr; weight, 73.4+/-16.6 kg; height, 167.6+/-9.7 cm; body mass index, 26.0+/-5.2 kg/m2) who had whole-body scans performed within a 3-h period on the 3 systems. Repeated measures ANOVA was used to test the null hypothesis that the mean values for the 3 systems were equal. Translation equations between the methods were derived using regression techniques.
RESULTS
Bone mineral content (BMC): For both genders, total BMC by iDXA was lower (p<or=0.004) than the other systems. Lean: for males, iDXA was lower (p<or=0.03) than the other systems for total, trunk, and arms. For females, DPXL estimated higher (p<0.001) lean mass compared with the other systems for total, trunk, and arms, but iDXA estimated greater legs lean mass. For both genders, all DPXL mean values were greater than Prodigy mean values (p<0.001). Fat: in females, all the 3 systems were different from each other for total, trunk, and legs (p<or=0.04). For arms, DPXL and iDXA were higher than Prodigy (p<0.0004). For males, DPXL was less (p<0.001) for total body, trunk, and legs compared with the other 2 systems and greater than Prodigy only for arms (p<0.0007). These data were used to derive translation equations between systems. For several measurements, the differences between systems were related to gender.
CONCLUSION
For estimation of BMC and body composition, there was high agreement between all DXA systems (R2=0.85-0.99). Even so, cross-calibration equations should be used to examine data across systems to avoid erroneous conclusions.
References
J Clin Endocrinol Metab. 1988 Oct;67(4):839-44
Appl Radiat Isot. 1998 May-Jun;49(5-6):515-6
Osteoporos Int. 2005 Dec;16(12):2157-67
Br J Radiol. 1993 Oct;66(790):902-6
Cited by 118 citations (total)
Comparative Study Evaluation Study Journal Article Research Support, Non-U.S. Gov't
A comparison of dual energy X-ray absorptiometry and bioelectrical impedance analysis to measure total and segmental body composition in healthy young adults.
Electronic publication date May 23, 2012
Siobhan Leahy [...] Philip Jakeman et al.
The aim of this study was to investigate the accuracy of BIA in the measurement of total body composition and...
Comparative Study Journal Article Randomized Controlled Trial Research Support, Non-U.S. Gov't
The effects of four hypocaloric diets containing different levels of sucrose or high fructose corn syrup on weight loss and related parameters.
Electronic publication date April 13, 2013
Joshua Lowndes [...] James M Rippe et al.
BACKGROUND
The replacement of sucrose with HFCS in food products has been suggested as playing a role in the development of...