Publications

Journal Article
Multicenter Study Observational Study Video-Audio Media
Ppublish 210 citations
Electronic publication date
August 31, 2017

View on Pubmed.gov 26995690 | View on doi.org 10.1016/j.gie.2...

Digital, single-operator cholangiopancreatoscopy in the diagnosis and management of pancreatobiliary disorders: a multicenter clinical experience (with video).

Search for publications in this journal Gastrointest Endosc - 84(4):649-55

Abstract

BACKGROUND AND AIMS

Digital cholangioscopes provide higher-resolution imaging of the pancreatobiliary tract compared with fiberoptic instruments. The role of a new, digital, single-operator cholangiopancreatoscopy (SOC) system for diagnosis and treatment of pancreatobiliary disorders in clinical practice is not known.

METHODS

We performed a multicenter, observational study of 105 consecutive patients with suspected pancreatobiliary disorders. The main outcome measures were (1) sensitivity and specificity of SOC visual appearance and biopsies in the diagnosis of indeterminate biliary strictures and (2) achieving complete duct clearance in patients with biliary or pancreatic duct stones.

RESULTS

A total of 98 cholangioscopy and 7 pancreatoscopy procedures were performed in 105 patients. Superior views of the ductal lumen and mucosa were obtained in all 44 patients with indeterminate biliary strictures. Among the 44 patients who underwent SOC-guided biopsies, the specimen was adequate for histologic evaluation in 43 patients (97.7%). The sensitivity and specificity of SOC visual impression for diagnosis of malignancy was 90% (95% confidence interval [CI], 69.9%-97.2%) and 95.8% (95% CI, 79.8%-99.3%), respectively. The sensitivity and specificity of SOC-guided biopsies for diagnosis of malignancy was 85% (95% CI, 64.0%-94.8%) and 100% (95% CI, 86.2%-100%). In patients with biliary or pancreatic duct stones (N = 36), complete duct clearance with stone removal in 1 session was accomplished in 86.1% of patients (31/36). Three patients (2.9%) experienced SOC-related adverse events that included cholangitis in 2 patients and postprocedure pancreatitis in 1 patient.

CONCLUSIONS

SOC has become an integral part of the ERCP armamentarium and has high accuracy in the evaluation of indeterminate biliary strictures. Complete stone clearance was achieved in all but 1 patient with challenging biliary or pancreatic duct stones. (

CLINICAL TRIAL REGISTRATION NUMBER

NCT01815619.).

References

No references provided.

Cited by 210 citations (total)

Journal Article

Safety and efficacy of SpyGlass cholangiopancreatoscopy in routine clinical practice in a regional Singapore hospital.

Electronic publication date July 24, 2020

Search for publications in this journal Singapore Med J

Tiing Leong Ang [...] Andrew Boon Eu Kwek et al.

INTRODUCTION

This study examined the efficacy and safety of cholangiopancreatoscopy via the SpyGlassâ„¢ system in routine clinical practice.

METHODS

The clinical data of...

Journal Article

Peroral cholangioscopy by SpyGlass DS versus CHF-B260 for evaluation of the lateral spread of extrahepatic cholangiocarcinoma.

Electronic publication date November 10, 2018

Search for publications in this journal Endosc Int Open

Yoshihide Kanno [...] Kei Ito et al.

A newly developed peroral cholangioscopy (POCS) system, SpyGlassDS has high maneuverability. This study aimed to evaluate acceptability of the accuracy...

Conflict of interest statement

No conflict of interest statement provided.

MeSH

Specialties

Copyright

Copyright © 2016 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.