View on Pubmed.gov 32108423 | View on doi.org 10.1002/ccd.288...
Use of the GORE® CARDIOFORM Septal Occluder for percutaneous closure of secundum atrial septal defects: Results of the multicenter U.S. IDE trial.
- The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
- Advocate Children's Hospital, Section of Pediatric Cardiology, University of Illinois Hospital, Chicago, Illinois.
- University of California, San Francisco, California.
- Children's Hospital of Michigan, Detroit, Michigan.
- Sanger Heart & Vascular Institute, Levine Children's Hospital, Charlotte, North Carolina.
GSO Investigator Group
Abstract
OBJECTIVE
To evaluate safety and efficacy of the GORE® CARDIOFORM Septal Occluder for percutaneous transcatheter closure of ostium secundum atrial septal defects.
BACKGROUND
The GORE® CARDIOFORM septal occluder is a double-disc, low profile, soft, conformable device, with distinct advantages over the GORE® HELEX® Septal Occluder.
METHODS
Subjects were enrolled in this single arm prospective study from 21 U.S. sites, and followed for 3 years. Primary endpoint was 6 month composite clinical success, comprised of technical success (implantation and retention of device), closure success (normalization of right heart size), no 30 day serious adverse events, and no device embolization or reintervention. Secondary endpoints included technical success, procedure success (technical success and ≤ 2 mm residual shunt at procedure conclusion), closure success (clinically insignificant or no residual shunt), and safety (freedom from 30 day serious adverse events and 6-month device events).
RESULTS
Between October 2012 and May 2015, 50 pivotal and 350 continued access subjects underwent attempted transcatheter GORE® CARDIOFORM Septal Occluder implantation. Median age was 6.9 years, and mean static defect diameter 9.7 ± 3.1 mm. Device placement was achieved in 93.5% (374/400). Composite clinical success was 90.2% and clinical closure success was 98.8% at 6 months. Freedom from serious adverse events was 98.3% at 30 days, with no device embolizations or reinterventions through 6 months.
CONCLUSIONS
The GORE® CARDIOFORM Septal Occluder has high composite clinical success and safety, performing well in defects ≤17 mm by stop flow stretched diameter. Single, multifenestrated, and deficient retroaortic tissue defects were well represented and successfully treated.
References
Webb G, Gatzoulis MA. Atrial septal defects in the adult: recent progress and overview. Circulation. 2006;114(15):1645-1653.
King TD, Thompson SL, Steiner C, Mills NL. Secundum atrial septal defect. Nonoperative closure during cardiac catheterization. JAMA. 1976;235(23):2506-2509.
Du ZD, Hijazi ZM, Kleinman CS, Silverman NH, Larntz K, Amplatzer I. Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol. 2002;39(11):1836-1844.
McElhinney DB, Quartermain MD, Kenny D, Alboliras E, Amin Z. Relative risk factors for cardiac erosion following Transcatheter closure of atrial Septal defects: a case-control study. Circulation. 2016;133(18):1738-1746.
Cited by 30 citations (total)
Journal Article Review
Outcomes of Device Closure of Atrial Septal Defects.
Electronic publication date August 29, 2020
P Syamasundar Rao
Several devices have been designed and tried over the years to percutaneously close atrial septal defects (ASDs). Most of the...
Case Reports Journal Article
Persistent Shunt After Closure of Ostium Secundum Atrial Septal Defect.
Electronic publication date January 31, 2025
Frank F Seghatol-Eslami [...] Michael H Lanier et al.
Conflict of interest statement
MeSH
- •
- •
- •
- •
- •
Specialties
- •
- •
- •
- •
Copyright
© 2020 Wiley Periodicals, Inc.