View on Pubmed.gov 34222960 | View on doi.org 10.1093/jacamr/...
The potential for developing new antimicrobial resistance from the use of medical devices containing chlorhexidine, minocycline, rifampicin and their combinations: a systematic review.
- Department of Infectious Diseases, Infection Control and Employee Health, University of Texas MD Anderson Cancer Center, Houston, TX, USA.
Abstract
BACKGROUND
Catheter infections remain one of the most persistent adverse events causing significant morbidity, economic impact and mortality. Several strategies have been proposed to reduce these infections including the use of catheters embedded with antibiotics and/or antiseptics. One reoccurring challenge is the fear that antimicrobial medical devices will induce resistance. The aim of this systematic review is to evaluate the evidence for induced antimicrobial resistance caused by exposure to antimicrobial medical devices.
METHODS
Four electronic databases [MEDLINE, Embase, Cumulative Index to Nursing and Allied Health Literature (CINAHL) and Scopus] were screened for studies published between 1983 and 2019 regarding assessment of microbial resistance with use of medical devices containing chlorhexidine, minocycline, rifampicin or combinations thereof. Development of new resistance, selection for tolerant organisms and 'no change in resistance' were assessed.
RESULTS
Forty-four publications, grouped by study type and stratified by drug assessed, were included for analyses. The majority of studies found no change in resistance after exposure to antimicrobial medical devices (13 , 2 , 20 clinical). Development of new resistance was commonly reported with the use of rifampicin as a single agent and only reported in one study assessing the minocycline/rifampicin combination (M/R); however, the increase in MIC was well below clinical relevance.
CONCLUSIONS
Emergence of new resistance to combinations of M/R, minocycline/rifampicin/chlorhexidine (M/R/CH) and chlorhexidine/silver sulfadiazine (CHXSS) was rare. No clinical trials confirmed its occurrence and some refuted it. The risk of development of new resistance to these antimicrobial combinations appears more fear-based than substantiated by clinical and experimental evidence but warrants continued surveillance.
References
Intensive Care Med. 2004 May;30(5):837-43
J Clin Oncol. 2004 Aug 1;22(15):3163-71
Chest. 2004 Nov;126(5):1628-35
Crit Care Med. 2011 Feb;39(2):245-51
Cited by 11 citations (total)
Journal Article
Chlorhexidine Mucoadhesive Buccal Tablets: The Impact of Formulation Design on Drug Delivery and Release Kinetics Using Conventional and Novel Dissolution Methods.
Electronic publication date June 3, 2021
Enas Al-Ani [...] Khalid Doudin et al.
Oropharyngeal candidiasis (OPC) is a mucosal infection caused by spp., and it is common among the immunocompromised. This condition...
Journal Article
Novel dalbavancin-PLLA implant coating prevents hematogenous infection in a minimally invasive mouse tail vein model.
Electronic publication date December 6, 2022
Marlen Kloss [...] Martina Sombetzki et al.
Infective/bacterial endocarditis is a rare but life-threatening disease with a hospital mortality rate of 22.7% and a 1-year mortality rate...
Conflict of interest statement
MeSH
- •
- •
- •
- •
- •
- •
Copyright
© The Author(s) 2020. Published by Oxford University Press on behalf of the British Society for Antimicrobial Chemotherapy.