Publications

Journal Article Preprint Epublish 3 citations

PubMed: 38559149 DOI: 10.1101/2024.03.08.583971

Transcranial Functional Ultrasound Imaging Detects Focused Ultrasound Neuromodulation Induced Hemodynamic Changes .

bioRxiv

Authors: Jonas Bendig, Christian Aurup, Samuel G Blackman, Erica P McCune, Seongyeon Kim, Elisa E Konofagou

Affiliations

  • Department of Biomedical Engineering, Columbia University, New York, NY, USA.
  • Department of Radiology, Columbia University, New York, NY, USA.
  • Department of Neurosurgery, Columbia University, New York, NY, USA.

Abstract

Background

Focused ultrasound (FUS) is an emerging non-invasive technique for neuromodulation in the central nervous system (CNS). Functional ultrasound imaging (fUSI) leverages ultrafast Power Doppler Imaging (PDI) to detect changes in cerebral blood volume (CBV), which correlate well with neuronal activity and thus hold promise to monitor brain responses to FUS.

Objective

Investigate the immediate and short-term effects of transcranial FUS neuromodulation in the brain with fUSI by characterizing hemodynamic responses.

Methods

We designed a setup that aligns a FUS transducer with a linear array to allow immediate subsequent monitoring of the hemodynamic response with fUSI during and after FUS neuromodulation (FUS-fUSI) in lightly anesthetized mice. We investigated the effects of varying pressures and transducer positions on the hemodynamic responses.

Results

We found that higher FUS pressures increase the size of the activated brain area, as well as the magnitude of change in CBV and could show that sham sonications did not produce hemodynamic responses. Unilateral sonications resulted in bilateral hemodynamic changes with a significantly stronger response on the ipsilateral side. FUS neuromodulation in mice with a cranial window showed distinct activation patterns that were frequency-dependent and different from the activation patterns observed in the transcranial model.

Conclusion

fUSI is hereby shown capable of transcranially monitoring online and short-term hemodynamic effects in the brain during and following FUS neuromodulation.

MeSH

Blood, Ultrasonography, Brain, Transducers, Blood Volume, Cerebral Blood Volume, Hemodynamics