Effective and organ doses assessment in neuroradiologic and hemodynamic procedures: comparison among three different simulation software.
Phys Med 136:105053
Authors: Francesca Calderoni, Andrea D'Alessio, Marco Maria Jacopo Felisi, Klarisa Elena Szilagyi, Cristina De Mattia, Mariangela Piano, Emanuela Piccaluga, Alessandro Turra, Marco Brambilla, Paola Enrica Colombo
Affiliations
- Medical Physics Department, University Hospital of Ferrara, Italy.
- Medical Physics Department, University Hospital Maggiore della Carità , Novara, Italy.
- Medical Physics Department, ASST Grande Ospedale Metropolitano Niguarda, Milano, Italy.
- Specialization School of Medical Physics, University of Bologna, Italy.
- Medical Physics Department, ASST Grande Ospedale Metropolitano Niguarda, Milano, Italy.
- Neuroradiology Department, ASST Grande Ospedale Metropolitano Niguarda, Milano, Italy.
- Hemodynamic Department, ASST Grande Ospedale Metropolitano Niguarda, Milano, Italy.
Abstract
INTRODUCTION
The goal of this work is to compare three different software for effective (ED) and organ (OD) dose estimation to analyze the reliability and investigate the main differences.
METHODS
Both thirty neuroradiologic and hemodynamic procedures were analyzed retrospectively choosing a representative sample in terms of kerma-area product and complexity. Each procedure was simulated using NCIRF, VirtualDose-IR Batch Utilityand PCXMC. The three software differ for computational method and phantom used and they also need different parameters from RDSR or experimental data to meet the requirements.Bland-Altman method was used to assess the agreement among software.
RESULTS
Large variations in some ODs were observed for all the software, ranging from -70Â % to 109Â % for neuroradiology and -73Â % to 92Â % for hemodynamics. This variation is greatly reduced when considering ED, from -23Â % to 19Â % and from -12Â % to 18Â % respectively. Bland-Altman analysis showed a poor agreement in OD estimations, with mean differences from 0.8 to 92Â mGy and variations up to 200Â mGy. The variability in ED is less evident, with mean differences from 1.3 to 3.7Â mSv for neuroradiology and from 0.6Â mSv to 3.4Â mSv for hemodynamics, although in individual cases these differences reached 7.6Â mSv and 11Â mSv respectively.
CONCLUSIONS
Differences principally arise from input parameters, models and phantoms which strongly affect ODs. Nevertheless, the overall EDs is confirmed to be a robust quantity with discrepancies lower than 4Â mSv (20Â %), showing differences comparable to uncertainties in radiological dosimetry and confirming its utility in population dose assessment.