View on Pubmed.gov 40730745 | View on doi.org 10.1007/s12194-...
Temporal changes in sucrose phantoms for magnetic resonance imaging: effects of preparation temperatures.
- Department of Molecular Imaging, Clinical Collaboration Unit, School of Medical Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-Cho, Toyoake, Aichi, 470-1192, Japan. yasuo.takatsu@fujita-hu.ac.jp.
- Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan. yasuo.takatsu@fujita-hu.ac.jp.
- Department of Radiology, Fujita Health University Hospital, 1-98 Dengakugakubo, Kutsukake-Cho, Toyoake, Aichi, 470-1192, Japan.
- Division of Medical Physics, School of Medical Sciences, Fujita Health University, 1-98 Dengakugakubo, Kutsukake-Cho, Toyoake, Aichi, 470-1192, Japan.
- Graduate School of Interdisciplinary Science and Engineering in Health Systems, Okayama University, 3-1-1 Tsushima-Naka, Kita-Ku, Okayama City, Okayama, 700-8530, Japan.
- Division of Regulatory Science, School of Medical Sciences, Fujita Health University, 1-98 Dengakugakubo, Kutsukake-Cho, Toyoake, Aichi, 470-1192, Japan.
- Major in Medical Sciences, Graduate School of Medical Sciences, Fujita Health University, 1-98 Dengakugakubo, Kutsukake-Cho, Toyoake, Aichi, 470-1192, Japan.
- Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
Abstract
This study investigated the effects of preparation temperature and usage period on the relaxation times and apparent diffusion coefficients (ADC) of sucrose phantoms to enhance imaging reliability. Phantoms were prepared using 10% sucrose solutions at 20, 50, and 80 °C. T1 and T2 relaxation times and ADC were monitored over 80 days using magnetic resonance imaging on a 1.5 T system. T2 relaxation time in 50 °C solutions increased from 245.5 to 1579 ms, while 80 °C solutions showed the highest stability (coefficient of variation ≈ 1.8%). T1 relaxation time changes were minimal, and ADC decreased at an average rate of 2.19 × 10 -6 mm 2/s per day. Bacteria were observed in the sucrose solution, and higher protein concentrations were strongly correlated with decreased 1/T2. In conclusion, sucrose phantoms exhibited temperature-dependent stability, with 80 °C preparations providing the most reliable T2 relaxation time.
References
Beck BL, Jenkins KA, Rocca JR, Fitzsimmons JR. Tissue-equivalent phantoms for high frequencies. Concepts Magn Reson Part B Magn Reson Eng. 2004;20B:30–3. doi.org .
Laubach HJ, Jakob PM, Loevblad KO, Baird AE, Bovo MP, Edelman RR, Warach S. A phantom for diffusion-weighted imaging of acute stroke. J Magn Reson Imaging. 1998;8:1349–54. doi.org .
Hara M, Kuroda M, Ohmura Y, Matsuzaki H, Kobayashi T, Murakami J, Katashima K, Ashida M, Ohno S, Asaumi J. A new phantom and empirical formula for apparent diffusion coefficient measurement by a 3 Tesla magnetic resonance imaging scanner. Oncol Lett. 2014;8:819–24. doi.org .
Winfield JM, Collins DJ, Priest AN, Quest RA, Glover A, Hunter S, Morgan VA, Freeman S, Rockall A, deSouza NM. A framework for optimization of diffusion-weighted MRI protocols for large field-of-view abdominal-pelvic imaging in multicenter studies. Med Phys. 2016;43:95. doi.org .
Conflict of interest statement
Declarations. Competing interests: The authors declare no conflicts of interest or competing interests relevant to the content of this article. Ethics approval: This study did not require approval from an ethics committee, because it was based on phantoms.
MeSH
- •
- •
- •
Specialties
- •
- •
- •
Copyright
© 2025. The Author(s), under exclusive licence to Japanese Society of Radiological Technology and Japan Society of Medical Physics.