Index of Device Efficiency: IVUS-Based Metric to Quantify Cutting Balloon Performance in Calcified Coronary Lesions.
- Department of Cardiology, Hôpital de Jolimont, La Louvrière, Belgium.
- University Hospital Basel, Basel, Switzerland.
- Department of Cardiology, Clinique St. Joseph-Groupe Vivalia, Arlon, Belgium.
- Cardiology Department, Ares Monza Hospital, Bucharest, Romania.
- Department of Cardiovascular Sciences, Campus Bio-medico University Hospital, Rome, Italy.
- Cardiology Department, "Niculae Stancioiu" Heart Institute, Cluj-Napoca, Romania.
- Department of Cardiovascular, Hôpital Valenciennes, Valenciennes, France.
- Cardio Center, IRCCS Humanitas Research Hospital, Rozzano-Milan, Italy.
- HartCentrum Ziekenhuis Aan de Stroom (ZAS) Middelheim, Antwerp, Belgium.
- St. George's University Hospital NHS Foundation Trust, London, UK.
- Department of Cardiology, McGovern School of Medicine, University of Texas, Austin, Texas, USA.
- Department of Cardiology, CHU Dupuytren, Limoges, France.
- Minneapolis Heart Institute and Minneapolis Heart Institute Foundation, Minneapolis, Minnesota, USA.
- Cardiology Department, CHU Saint Etienne, Saint-Priest-en-Jarez, France.
Abstract
BACKGROUND
Calcified coronary artery disease complicates percutaneous coronary intervention (PCI), yet no standardized parameter quantifies the efficiency of plaque-modifying devices.
AIMS
To introduce and evaluate the Index of Device Efficiency (IDE), a novel intravascular ultrasound (IVUS)-derived metric quantifying the mechanical efficiency of cutting balloon (CB) angioplasty during the RODIN-CUT technique.
METHODS
In this multicenter retrospective study, 50 heavily calcified lesions (47 patients) underwent 270 IVUS pullbacks. The IDE was defined as the ratio of maximal luminal diameter achieved to the nominal CB diameter ×100. Stepwise IDE dynamics were assessed after sequential high-pressure CB inflations. A plateau phase was defined as ≤ 0.3 mm further luminal diameter gain or ≤ 0.5 mm² area gain between inflations. The relationship between IDE ≥ 100% and post-stent expansion was evaluated. Safety outcomes were systematically recorded.
RESULTS
Mean age was 75 ± 8 years; 46% of lesions were in the left anterior descending artery and 30% in the left main bifurcation. Minimal lumen area increased from 4.0 ± 1.5 mm² at baseline to 5.8 ± 2.1 mm² after one inflation and to 8.8 ± 2.6 mm² after four inflations (p < 0.001). IDE ≥ 100% was reached in 40% of lesions after one inflation versus 92% after four, with no further gain thereafter (plateau). Lesions with IDE ≥ 100% more frequently achieved optimal stent expansion (82% vs. 10%, p = 0.007). No major procedural complications occurred.
CONCLUSIONS
IDE is a simple, reproducible IVUS-derived metric that quantifies CB efficiency, identifies a reproducible plateau after four inflations, and predicts optimal stent expansion.
References
R. F. Riley, M. P. Patel, J. D. Abbott, et al., “SCAI Expert Consensus Statement on the Management of Calcified Coronary Lesions,” Journal of the Society for Cardiovascular Angiography & Interventions 3, no. 2 (2024): 101259, doi.org
M. Alkhouli, “Coronary Calcifications in Patients Undergoing PCI,” JACC: Cardiovascular Interventions 13, no. 12 (2020): 1429–1431.
G. Souteyrand, T. Mouyen, B. Honton, et al., “Stent Underexpansion Is an Underestimated Cause of Intrastent Restenosis: Insights From RESTO Registry,” Journal of the American Heart Association 13, no. 21 (2024): e036065.
M. Shah, O. Najam, R. Bhindi, and K. De Silva, “Calcium Modification Techniques in Complex Percutaneous Coronary Intervention,” Circulation: Cardiovascular Interventions 14, no. 5 (2021): E009870.