The value of microelectrodes and microbipolar electrograms in premature ventricular complex mapping and ablation: The micro-PVC study.

Search for publications in this journal Heart Rhythm O2 - 7(6):1157-1165

Abstract

BACKGROUND

Catheter ablation is an effective treatment for premature ventricular contractions (PVCs). Mapping typically relies on bipolar (bip-EGMs) and unipolar electrograms (uni-EGMs). Recently, catheters equipped with microbipolar electrodes have become available.

OBJECTIVE

This study aimed to investigate the value of microbipolar EGMs (micro-EGMs) in PVCs mapping.

METHODS

We performed a multicenter retrospective study including patients who underwent successful outflow tract PVC ablation using QDOT catheters. 2 groups were identified based on the superficial or intramural origin of the PVC. EGMs-bip-, uni-, and micro-EGMs-were assessed at the successful ablation sites. The primary outcome was the ability of micro-EGM to discriminate between superficial and intramural PVCs. The secondary outcome was signal amplitude.

RESULTS

Among 81 patients (mean age 54 ± 11 years; 59.3% male), 51 had superficial and 30 had intramural PVCs. In superficial PVC, micro-EGM preceded both uni-EGM and bip-EGM in 90% of cases. In particular, when the earliest bip-EGM preceded the earliest micro-EGM by >10 ms (Δ bip-micro >10 ms), there was a high probability (sensitivity 0.94; specificity 0.75) of requiring ablation in both the right ventricular outflow tract and left ventricular outflow tract for successful PVC suppression. Micro-EGMs had higher initial deflection amplitude (1.5 vs 2.8 mV; P = .025).

CONCLUSION

Micro-EGMs allow more accurate PVC mapping, improving depth estimation. Micro-EGMs seem to be valuable for local activation time annotation with important implications on automated annotation algorithms within existing electroanatomic mapping systems.

References

Enriquez A., Muser D., Markman T.M., Garcia F. Mapping and ablation of premature ventricular complexes: state of the art. JACC Clin Electrophysiol. 2024;10:1206–1222.

Stevenson W.G., Soejima K. Recording techniques for clinical electrophysiology. J Cardiovasc Electrophysiol. 2005;16:1017–1022.

El H.M., Houben R., Stroobandt R., Van H.F., Tavernier R., Duytschaever M. Novel algorithmic methods in mapping of atrial and ventricular tachycardia. Circ Arrhythm Electrophysiol. 2014;7:463–472.

Higuchi K., Yavin H.D., Sroubek J., Younis A., Zilberman I., Anter E. How to use bipolar and unipolar electrograms for selecting successful ablation sites of ventricular premature contractions. Heart Rhythm. 2022;19:1067–1073.

Conflict of interest statement

M.B. reports receiving speaker honoraria from Johnson & Johnson, consulting fees from Boston Scientific, and a research grant from Johnson & Johnson and Boston Scientific. N.S. reports consulting fees from Boston Scientific, Abbott, and Johnson & Johnson. The authors declare that all illustrations and figures in the manuscript are entirely original and do not require reprint permission. The authors have no conflicts of interest to disclose.

MeSH

Specialties

Copyright

© 2026 Heart Rhythm Society. Published by Elsevier Inc.