Publications

Comparative Study
Journal Article
Ppublish 463 citations
Electronic publication date
July 1, 1994

View on Pubmed.gov 8013092 View on doi.org

Limit of detection (LQD)/limit of quantitation (LOQ): comparison of the empirical and the statistical methods exemplified with GC-MS assays of abused drugs.

Search for publications in this journal Clin Chem - 40(7 Pt 1):1233-8

  • D A Armbruster

    Search for this expert
    1
  • M D Tillman

    Search for this expert
  • L M Hubbs

    Search for this expert

Abstract

The limit of detection (LOD) for any analytical procedure, the point at which analysis is just feasible, may be determined by a statistical approach based on measuring replicate blank (negative) samples or by an empirical approach, consisting of measuring progressively more dilute concentrations of analyte. The limit of quantitation (LOQ), or concentration at which quantitative results can be reported with a high degree of confidence, may likewise be determined by either approach. We used both methods to determine LOD and LOQ for forensic gas chromatographic-mass spectrometric (GC-MS) analyses of abused drugs. The statistically determined LOD and LOQ values for these assays underestimated the LOD because of the large imprecision associated with blank measurements and the inability of blank samples to meet typical GC-MS acceptance criteria. The empirical method provided much more realistic LOD values, supported by reasonable experimental data, and are 0.5-0.03 the magnitude of the corresponding statistical LODs. The empirical LODs and LOQs are identical for these GC-MS assays. The observations made here about the LOD/LOQ for specific forensic GC-MS procedures are generally applicable to any type of analysis.

References

No references provided.

Cited by 463 citations (total)

Journal Article Research Support, N.I.H., Extramural Research Support, Non-U.S. Gov't Research Support, U.S. Gov't, Non-P.H.S.

Quantitative determination of paraquat in meconium by sodium borohydride-nickel chloride chemical reduction and gas chromatography/mass spectrometry (GC/MS).

Electronic publication date April 2, 2008

Search for publications in this journal J Chromatogr B Analyt Technol Biomed Life Sci

Norberto C Posecion [...] Dawn M Bielawski et al.

The objective of this study was to develop a procedure for the GC/MS assay of paraquat in meconium as a...

Journal Article Research Support, N.I.H., Extramural

Change-point models to estimate the limit of detection.

Electronic publication date October 28, 2014

Search for publications in this journal Stat Med

Ryan C May [...] David M Margolis et al.

In many biological and environmental studies, measured data is subject to a limit of detection. The limit of detection is...

Conflict of interest statement

No conflict of interest statement provided.

MeSH