Publications

Comparative Study
Journal Article
Ppublish 122 citations
Electronic publication date
April 9, 2011

View on Pubmed.gov 20817019 | View on doi.org 10.1016/j.hrthm...

Cryoballoon versus radiofrequency catheter ablation of paroxysmal atrial fibrillation: biomarkers of myocardial injury, recurrence rates, and pulmonary vein reconnection patterns.

Search for publications in this journal Heart Rhythm - 7(12):1770-6

Abstract

BACKGROUND

Cryoballoon ablation has emerged as a novel treatment strategy for patients with atrial fibrillation (AF).

OBJECTIVE

The purpose of this study was to compare pulmonary vein isolation (PVI) using cryoballoon ablation versus RF ablation with regard to myocardial injury, pulmonary vein (PV) reconnection patterns, and outcome.

METHODS

Fifty patients (age 59 ± 9 years, ejection fraction 0.59 ± 0.06, left atrial size 41 ± 5 mm) with paroxysmal AF were studied. Twenty-five patients underwent PVI using a 28-mm cryoballoon. A control group of 25 patients underwent PVI using an open-irrigation RF ablation catheter. Myocardial injury was determined by measuring troponin T (TnT). PV reconnection patterns were studied in case of repeat procedures.

RESULTS

Procedure duration was 166 ± 32 minutes in the cryoballoon group versus 197 ± 52 minutes in the RF group (P = .014), with similar ablation times (cryoballoon: 45 minutes [interquartile range 40-52.5 minutes]; RF: 47 minutes [interquartile range 44-65 minutes], P = .17). Postprocedural TnT in the RF group was 1.29 ± 0.41 μg/L versus 0.76 ± 0.55 μg/L in the cryoballoon group (P = .002). In 12 patients who underwent repeat ablation, 74% of PV reconnection sites were inferiorly located in the cryoballoon group compared to 17% in the RF group (P = .0004). With 1.2 ± 0.4 and 1.3 ± 0.6 procedures per patient, 88% of patients in the cryoballoon group and 92% in the RF group were in stable sinus rhythm after follow-up of 12 ± 3 months (P = NS).

CONCLUSION

Differences in the extent of myocardial injury and patterns of PV reconnection were observed between cryoballoon ablation and RF ablation of paroxysmal AF.

References

No references provided.

Cited by 122 citations (total)

Comparative Study Controlled Clinical Trial Journal Article

Radiofrequency catheter ablation versus balloon cryoablation of atrial fibrillation: markers of myocardial damage, inflammation, and thrombogenesis.

Electronic publication date May 6, 2017

Search for publications in this journal Wien Klin Wochenschr

Bor Antolič [...] Matjaž Šinkovec et al.

BACKGROUND

Evidence from animal and human studies suggests that cryoablation might be associated with a lesser inflammatory response and activation of...

Journal Article Review

Treating Atrial Fibrillation With Cryoballoon Technology.

Electronic publication date April 14, 2012

Search for publications in this journal J Atr Fibrillation

David R Altmann [...] Michael Kühne et al.

Cryoballoon ablation has emerged as a novel tool to perform pulmonary vein isolation. The aim of this paper is to...