Resonant Mode Reduction in Radiofrequency Volume Coils for Ultrahigh Field Magnetic Resonance Imaging.

Search for publications in this journal Materials (Basel) - 4(8):1333-1344

Abstract

In a multimodal volume coil, only one mode can generate homogeneous Radiofrequency (RF) field for Magnetic Resonance Imaging. The existence of other modes may increase the volume coil design difficulties and potentially decreases coil performance. In this study, we introduce common-mode resonator technique to high and ultrahigh field volume coil designs to reduce the resonant mode while maintain the homogeneity of the RF field. To investigate the design method, the common-mode resonator was realized by using a microstrip line which was split along the central to become a pair of parallel transmission lines within which common-mode currents exist. Eight common-mode resonators were placed equidistantly along the circumference of a low loss dielectric cylinder to form a volume coil. Theoretical analysis and comparison between the 16-strut common-mode volume coil and a conventional 16-strut volume coil in terms of RF field homogeneity and efficiency was performed using Finite-Difference Time-Domain (FDTD) method at 298.2 MHz. MR imaging experiments were performed by using a prototype of the common-mode volume coil on a whole body 7 Tesla scanner. FDTD simulation results showed the reduced number of resonant modes of the common-mode volume coil over the conventional volume coil, while the RF field homogeneity of the two type volume coils was kept at the same level. MR imaging of a water phantom and a kiwi fruit showing the feasibility of the proposed method for simplifying the volume coil design is also presented.

References

J Magn Reson Imaging. 2009 Nov;30(5):1209-14

IEEE Trans Med Imaging. 2010 Jan;29(1):179-84

AJNR Am J Neuroradiol. 2001 Nov-Dec;22(10):1881-6

IEEE Trans Med Imaging. 1989;8(3):286-94

Cited by 18 citations (total)

Journal Article Research Support, N.I.H., Extramural Research Support, Non-U.S. Gov't

A dual-tuned quadrature volume coil with mixed λ/2 and λ/4 microstrip resonators for multinuclear MRSI at 7 T.

Electronic publication date May 18, 2012

Search for publications in this journal Magn Reson Imaging

Yong Pang [...] Xiaoliang Zhang et al.

In this work, an eight-element by eight-element dual-tuned quadrature volume coil with a mix of capacitor terminated half-wavelength (λ/2) and...

Preprint

Electric Field and SAR Reduction in High Impedance RF Arrays by Using High Permittivity Materials for 7T MR Imaging.

Electronic publication date December 18, 2023

Search for publications in this journal ArXiv

Aditya A Bhosale [...] Xiaoliang Zhang et al.

Higher frequencies and shorter wavelengths present significant design issues at ultra-high fields, making multi-channel array setup a critical component for...

Conflict of interest statement

No conflict of interest statement provided.

MeSH

Specialties