An RF dosimeter for independent SAR measurement in MRI scanners.

Search for publications in this journal Med Phys - 40(12):122303

Abstract

PURPOSE

The monitoring and management of radio frequency (RF) exposure is critical for ensuring magnetic resonance imaging (MRI) safety. Commercial MRI scanners can overestimate specific absorption rates (SAR) and improperly restrict clinical MRI scans or the application of new MRI sequences, while underestimation of SAR can lead to tissue heating and thermal injury. Accurate scanner-independent RF dosimetry is essential for measuring actual exposure when SAR is critical for ensuring regulatory compliance and MRI safety, for establishing RF exposure while evaluating interventional leads and devices, and for routine MRI quality assessment by medical physicists. However, at present there are no scanner-independent SAR dosimeters.

METHODS

An SAR dosimeter with an RF transducer comprises two orthogonal, rectangular copper loops and a spherical MRI phantom. The transducer is placed in the magnet bore and calibrated to approximate the resistive loading of the scanner's whole-body birdcage RF coil for human subjects in Philips, GE and Siemens 3 tesla (3T) MRI scanners. The transducer loop reactances are adjusted to minimize interference with the transmit RF field (B1) at the MRI frequency. Power from the RF transducer is sampled with a high dynamic range power monitor and recorded on a computer. The deposited power is calibrated and tested on eight different MRI scanners. Whole-body absorbed power vs weight and body mass index (BMI) is measured directly on 26 subjects.

RESULTS

A single linear calibration curve sufficed for RF dosimetry at 127.8 MHz on three different Philips and three GE 3T MRI scanners. An RF dosimeter operating at 123.2 MHz on two Siemens 3T scanners required a separate transducer and a slightly different calibration curve. Measurement accuracy was ∼3%. With the torso landmarked at the xiphoid, human adult whole-body absorbed power varied approximately linearly with patient weight and BMI. This indicates that whole-body torso SAR is on average independent of the imaging subject, albeit with fluctuations.

CONCLUSIONS

Our 3T RF dosimeter and transducers accurately measure RF exposure in body-equivalent loads and provide scanner-independent assessments of whole-body RF power deposition for establishing safety compliance useful for MRI sequence and device testing.

References

J Magn Reson Imaging. 2000 Jul;12(1):30-6

Phys Med Biol. 2001 Feb;46(2):609-19

Magn Reson Med. 2001 Jul;46(1):24-30

J Magn Reson Imaging. 2001 Oct;14(4):488-9

Cited by 21 citations (total)

Journal Article

Use of a radio frequency shield during 1.5 and 3.0 Tesla magnetic resonance imaging: experimental evaluation.

Electronic publication date November 8, 2014

Search for publications in this journal Med Devices (Auckl)

Christopher P Favazza [...] Krzysztof R Gorny et al.

Radiofrequency (RF) shields have been recently developed for the purpose of shielding portions of the patient's body during magnetic resonance...

Journal Article

MRI scanner-independent specific absorption rate measurements using diffusion coefficients.

Electronic publication date April 3, 2018

Search for publications in this journal J Appl Clin Med Phys

Youngseob Seo [...] Zhiyue J Wang et al.

OBJECTIVE

The purpose of this study was to measure specific absorption rate (SAR) during MRI scanning using a human torso phantom...