B1 field homogeneity and SAR calculations for the birdcage coil.
- Department of Electrical Engineering, The Ohio State University, Columbus 43212, USA.
Abstract
The finite-difference time-domain (FDTD) method is used to model a birdcage resonator. All the coil components, including the wires, lumped capacitors and the source, are geometrically modelled together. As such, the coupling effects within the birdcage, including the interactions of coil, source and human head, are accurately computed. A study of the transverse magnetic (B1) field homogeneity and the specific absorption rate (SAR) is presented on an anatomically detailed human head model at 64 and 200 MHz representing 1.5 and 4.7 T MRI systems respectively. Unlike that at 64 MHz, the B1 field distribution is found to be inhomogeneous at 200 MHz. Also, high local SAR values are observed in the tissue near the source due to the coupling between the source and the head at 200 MHz.
References
Cited by 80 citations (total)
Journal Article
Resonant Mode Reduction in Radiofrequency Volume Coils for Ultrahigh Field Magnetic Resonance Imaging.
Electronic publication date November 15, 2011
Yong Pang [...] Xiaoliang Zhang et al.
In a multimodal volume coil, only one mode can generate homogeneous Radiofrequency (RF) field for Magnetic Resonance Imaging. The existence...
Journal Article Research Support, N.I.H., Extramural
An RF dosimeter for independent SAR measurement in MRI scanners.
Electronic publication date August 7, 2014
Di Qian [...] William A Edelstein et al.
PURPOSE
The monitoring and management of radio frequency (RF) exposure is critical for ensuring magnetic resonance imaging (MRI) safety. Commercial MRI...