Optimization of pseudo-continuous arterial spin labeling using off-resonance compensation strategies at 7T.
Magn Reson Med 87(4):1720-1730
Authors: Gaël Saïb, Alan P Koretsky, S Lalith Talagala
Affiliations
- NINDS/LFMI, National Institutes of Health, Bethesda, Maryland, USA.
- NINDS/NMRF, National Institutes of Health, Bethesda, Maryland, USA.
Abstract
PURPOSE
The sensitivity of pseudo-continuous arterial spin labeling (PCASL) to off-resonance effects (ΔB 0 ) is a major limitation at ultra-high field (≥7T). The aim of this study was to assess the effectiveness of different PCASL ΔB 0 compensation methods at 7T and measure the labeling efficiency with off-resonance correction.
THEORY AND METHODS
Phase offset errors induced by ΔB 0 at the feeding arteries can be compensated by adding an extra radiofrequency (RF) phase increment and transverse gradient blips into the PCASL RF pulse train. The effectiveness of an average field correction (AVGcor), a vessel-specific field-map-based correction (FMcor) and a vessel-specific prescan-based correction (PScor) were compared at 7T. After correction, the PCASL labeling efficiency was directly measured in feeding arteries downstream from the labeling location.
RESULTS
The perfusion signal was more uniform throughout the brain after off-resonance correction. Whole-brain average perfusion signal increased by a factor of 2.4, 2.5, and 2.1, respectively, with AVGcor, FMcor and PScor compared to acquisitions without correction. With off-resonance correction, the maximum labeling efficiency was ~0.68 at mean B 1 (B 1mean ) of 0.70 µT when using a mean gradient (G mean ) of 0.25 mT/m.
CONCLUSION
Either a prescan or a field map can be used to correct for off-resonance effects and retrieve a good brain perfusion signal at 7T. Although the three methods performed well in this study, FMcor may be better suited for patient studies because it accounted for vessel-specific ΔB 0 variations. Further improvements in image quality will be possible by optimizing the labeling efficiency with advanced hardware and software while satisfying specific absorption rate constraints.
MeSH
Brain, Arteries
Cited by
- Whole-cerebrum three-dimensional pseudo-continuous arterial spin labeling at 5T: reproducibility and preliminary application in moyamoya. Quant Imaging Med Surg, 2025
- Dynamic B field shimming for improving pseudo-continuous arterial spin labeling at 7 T. Magn Reson Med, 2025
- Optimization of pseudo-continuous arterial spin labeling using off-resonance compensation strategies at 7T. Magn Reson Med, 2022
- Whole-Cerebrum distortion-free three-dimensional pseudo-Continuous Arterial Spin Labeling at 7T. medRxiv, 2023
- SNR-efficient whole-brain pseudo-continuous arterial spin labeling perfusion imaging at 7 T. Magn Reson Med, 2025