PubMed: 36255158 DOI: 10.1002/mrm.29491
Time-encoded pseudo-continuous arterial spin labeling: Increasing SNR in ASL dynamic angiography.
Magn Reson Med 89(4):1323-1341
Authors: Joseph G Woods, S Sophie Schauman, Mark Chiew, Michael A Chappell, Thomas W Okell
Affiliations
- Wellcome Centre for Integrated Neuroimaging, FMRIB, Nuffield Department of Clinical Neuroscience, University of Oxford, Oxford, United Kingdom.
- Department of Radiology, Stanford University, Stanford, California, USA.
- Mental Health and Clinical Neurosciences, School of Medicine, University of Nottingham, Nottingham, United Kingdom.
- Sir Peter Mansfield Imaging Centre, School of Medicine, University of Nottingham, Nottingham, United Kingdom.
- Nottingham Biomedical Research Centre, Queen's Medical Centre, University of Nottingham, Nottingham, United Kingdom.
Abstract
PURPOSE
Dynamic angiography using arterial spin labeling (ASL) can provide detailed hemodynamic information. However, the long time-resolved readouts require small flip angles to preserve ASL signal for later timepoints, limiting SNR. By using time-encoded ASL to generate temporal information, the readout can be shortened. Here, the SNR improvements from using larger flip angles, made possible by the shorter readout, are quantitatively investigated.
METHODS
The SNR of a conventional protocol with nine Look-Locker readouts and a 4 3 time-encoded protocol with three Look-Locker readouts (giving nine matched timepoints) were compared using simulations and in vivo data. Both protocols were compared using readouts with constant flip angles (CFAs) and variable flip angles (VFAs), where the VFA scheme was designed to produce a consistent ASL signal across readouts. Optimization of the background suppression to minimize physiological noise across readouts was also explored.
RESULTS
The time-encoded protocol increased in vivo SNR by 103% and 96% when using CFAs or VFAs, respectively. Use of VFAs improved SNR compared with CFAs by 25% and 21% for the conventional and time-encoded protocols, respectively. The VFA scheme also removed signal discontinuities in the time-encoded data. Preliminary data suggest that optimizing the background suppression could improve in vivo SNR by a further 16%.
CONCLUSIONS
Time encoding can be used to generate additional temporal information in ASL angiography. This enables the use of larger flip angles, which can double the SNR compared with a non-time-encoded protocol. The shortened time-encoded readout can also lead to improved background suppression, reducing physiological noise and further improving SNR.
MeSH
Angiography, Preliminary Data
Cited by
- Optimization of 4D combined angiography and perfusion using radial imaging and arterial spin labeling. Magn Reson Med, 2023
- Time-encoded pseudo-continuous arterial spin labeling: Increasing SNR in ASL dynamic angiography. Magn Reson Med, 2023
- Transformer-based deep learning denoising of single and multi-delay 3D arterial spin labeling. Magn Reson Med, 2023
- Ultra-high temporal resolution 4D angiography using arterial spin labeling with subspace reconstruction. Magn Reson Med, 2025