Publications

Journal Article Ahead of print

PubMed: 41610593 DOI: 10.1016/j.ultramic.2026.114324

Enhancing atomic force microscopy stability through second harmonic optical fibre cavity control.

Ultramicroscopy 282:114324

Authors: Joana Nobre, Tiago Cordeiro, Tiago T Robalo, Ana P Carapeto, Mário S Rodrigues

Affiliations

  • Departamento de Física, Faculdade de Ciências, Universidade de Lisboa, Campo Grande 16, 1749-016 Lisbon, Portugal; BioISI-Biosystems and Integrative Sciences Institute, Faculdade de Ciências, Universidade de Lisboa, Campo Grande 16, 1749-016 Lisbon, Portugal.
  • Departamento de Física, Faculdade de Ciências, Universidade de Lisboa, Campo Grande 16, 1749-016 Lisbon, Portugal; BioISI-Biosystems and Integrative Sciences Institute, Faculdade de Ciências, Universidade de Lisboa, Campo Grande 16, 1749-016 Lisbon, Portugal.

Abstract

The interatomic potential is key to understanding all properties of materials. Yet, conventional atomic force microscopes do not usually measure the interatomic potential, because they lack control of the tip-sample distance. Here, we propose a simple methodology for measuring tip-surface interactions directly as a function of tip-sample distance, rather than deflection as a function of sample displacement. We use an AC interferometer to monitor the absolute tip displacement. When a force is applied on the tip a negative feedback loop displaces the cantilever anchoring point, deflecting the lever such that the tip position remains constant. This feedback loop actively maintains the tip at a distance from an optical fibre where the interferometer sensitivity is maximum. As a result, the tip-sample distance or the indentation is directly given by the sample motion and the tip does not jump-to-contact.

MeSH

Feedback, Displacement, Psychological, Microscopy, Atomic Force, Microscopy, Electron, Microscopy