Navigating Porous Threaded Implants Across the SI Joint is Associated with a Lower Reported Rate of Symptomatic Acute Malposition.

Search for publications in this journal Med Devices (Auckl)

Abstract

OBJECTIVE

To determine whether use of intraoperative navigation is associated with a reduced rate of acute symptomatic implant malposition during iliosacral implant placement in sacroiliac (SI) joint fusion and sacral fracture fixation.

METHODS

Retrospective analysis of post-market databases regarding device use and reported device-related complaints from a single device manufacturer between March 2021 and August 2026 were combined, including procedure type, procedure date and use of navigation systems. Analysis focused on lateral and posterolateral (PL) transiliac placement of a porous threaded implant (PTI, iFuse TORQ) during SI joint fusion and sacral fracture fixation procedures performed in the USA. All acute symptomatic implant malpositions reported as product complaints were identified and linked to index surgeries through manual identification. The proportion of navigated and non-navigated cases that resulted in symptomatic implant malposition was compared using Firth's penalized logistic regression test.

RESULTS

25,096 procedures using 68,352 PTI were identified. Navigation was used in 15.6% (n=3,919) of cases. Implant malposition causing acute symptoms occurred in 63 cases (overall rate 0.25%). Of these, 2 occurred in navigated cases (rate, 0.05%) and 61 occurred in non-navigated cases (rate, 0.29%, odds ratio 0.22, 95% CI 0.046-0.639, p=0.0014).

CONCLUSION

Navigation was associated with a markedly lower rate of acute symptomatic implant malposition during lateral or PL transiliac placement of a PTI.

References

No references provided.

Conflict of interest statement

Two study authors (Joshi, Cher) work for SI-BONE, Inc., the manufacturer of the device under study. Two authors (Rajpal and Polly) are consultants to SI-BONE. Dr Sharad Rajpal reports Royalties or licenses from SI-BONE; Consulting fees from Medtronic, SI-BONE, Cerapedics and Zimmer-Biomet; Honoraria from Medtronic; Patents from SI-BONE, outside the submitted work. Dr David Polly reports Grants or contracts from Curiteva, Spinal Stabilization Technologies, Medtronic, SI-BONE; Royalties or licenses from SI-BONE; Consulting fees from Curiteva, Spinal Stabilization Technologies, Medtronic, SI-BONE; Patents from SI-BONEe; Data safety/advisory board participation from SI-BONE; Leadership or fiduciary roles from Minnesota Orthopedic Society; Stock or stock options from See All AI; Equipment/materials/services from SI-BONE, Medtronic, outside the submitted work. The authors report no other conflicts of interest in this work.